Error in spectrum window

 
Post new topic   Reply to topic    Pico Technology Forum Index -> PicoScope
Author Message
Guest






PostPosted: Mon Jan 21, 2008 9:07 pm    Post subject: Error in spectrum window Reply with quote

When doing spectrum analyze on a sine wave I get different level depending on the setting on the bandwith. Very strange. I'm talking about 5-10% or so.

Is this a known problem?

I'm running the latest (I think) Picoscope6.

Thanks in advance!


/Peter
Back to top
PeterF
Site Admin
Site Admin


Joined: 07 Jun 2007
Posts: 117
Location: Cambridgeshire

PostPosted: Tue Jan 22, 2008 11:28 am    Post subject: Reply with quote

Hi,
There will be some variation when the bandwidth is changed. The most accurate readings will be when the measured frequency is just below the upper range of measurement (a full scale reading). At higher bandwidth settings, when the peak is only a fraction of the full bandwidth, accuracy will suffer. The number of frequency bins selected will also have an effect.
PeterF.
Back to top
Audioguy
Super User
Super User


Joined: 09 Aug 2007
Posts: 22

PostPosted: Wed Feb 06, 2008 11:12 am    Post subject: Reply with quote

Thanks!

I was aware of the relationsship between FFT points and frequency of sampling.. must have had an off day. Smile


/Peter
Back to top