| Author |
Message |
Jiangping Yu Guest
|
Posted: Fri Feb 10, 2006 8:02 pm Post subject: PicoLog for windows ADC16: wrong response |
|
|
We have ADC 16 and picolog for windows software. During recording, the data will changed suddenly from 0.3 to -1.5 and keep at -1.5 without any change though the panel display is 0.3.
I want to know if the PicoLog for windows software caused this problem or ADC 16 connection has problem.
Thank you very much!
Jiangping |
|
| Back to top |
|
 |
Sarah Guest
|
Posted: Tue Feb 14, 2006 1:23 pm Post subject: |
|
|
Hi
Thank you for your post.
It could be the voltages on the RS232. If they are not supplying enough voltage then it could fail to respond intermittently. Check that the following is true on the serial port:
Pin 4 = -7V
Pin 5 = GND
Pin 7 = +7V
If this is the case and the problem persists then let me know.
Best Regards |
|
| Back to top |
|
 |
Jiangping Yu Guest
|
Posted: Tue Feb 14, 2006 9:17 pm Post subject: ADC16, wrong response |
|
|
Problem: PicoLog data suddenly dropped from 0.3 to -1.5 though the panel display is 0.3. I have to power off the machine and shut down picoLog, then restart to solve this problem. It happens very often and disturbs my normal tests.
You mentioned if the voltages on the RS232 is not supplying enough voltage then it could fail to respond intermittently.
There are 25pin connector on ADC16 panel. Pin 1: -5 v, Pin 2: +5
and from Pin6 to pin 13 are channels 8 to 1. Is this input voltage or supplying voltage for RS232?
RS232 has 9 pins:
RS232
o o o o
o o o o o
Could you please tell me which pins are 4, 5, 7? You said Pin 4=-7v,pin 5=GND, Pin 7= +7 v. Can I use muiltimeter to measure them? And how?
Thank you very much. |
|
| Back to top |
|
 |
Michael Guru

Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire
|
Posted: Wed Feb 22, 2006 9:56 am Post subject: |
|
|
Hello,
Reply sent by email.
Regards,
Michael |
|
| Back to top |
|
 |
|
Powered by phpBB © 2001, 2005 phpBB Group
|