The ADC-16 is actually quite slow in comparison to our other loggers and may not be fast enough for your application. The fastest it can sample on a single channel is 3 samples every two seconds at full resolution.
The ADC-16's conversion time is affected by the resolution that you select to use:-
Resolution (bits) Conversion Time(ms)
8 ....................6.6
9 ....................8.9
10 ..................14
11 ..................23
12 ..................41
13 ..................78
14 ..................151
15 ..................298
16 ..................657
Please contact on Helpdesk if you want help choosing another data logger:-