I'm thinking of buying a 3204 scope and I got two questions :

1. is the ground/shield of the probes connected to the ground of the USB, or is there a galvanic isolation?

2. Is it right that in the highest accuracy of the scope (100mV/screen with 8-bit resolution) the smallest horizontal step corresponds to approx. 0.4 mV?

As an explanation : I want to use the scope for current measurement, by measuring the voltage drop over a shunt (0,2 Ohm). Therefore I need to make a differential measurement over the shunt (by connecting the shield of the probe to one side of the shunt).
Back to top
ziko
Site Admin
Site Admin


Joined: 01 Dec 2006
Posts: 600
Location: St Neots

PostPosted: Mon Mar 26, 2007 2:40 pm    Post subject: Reply with quote

Hi and thank you for your post.

1) Yes they are connected and not there is no isolation
2)No, the smallest resolution would be 0.8mV , you forgot the sign bit.

Hope this helps.