| Author |
Message |
RobH Active User

Joined: 01 Nov 2006
Posts: 6
|
Posted: Wed Nov 01, 2006 3:13 pm Post subject: ADC-16 Charge & Current measurements |
|
|
Hi
Background:
I'm using and ADC-16 to measure a voltage drop across a low value resistor (0.25 ohms). This resistor is in series along the return line from a product to its battery.
I use a calculated parameter to show the current, log that for around 1/2 an hour, then (using excel) integrate current over time intervals to get instantaneous charge, sum the charges and divide by total time to get average current. (Simple, eh?!?!)
Problem:
I need to measure the 'off' (or Standby) current of the product, typically nano-amps, but the drop across the resistor is not enough to be seen by the logger.
Attempted solutions:
A previous solution was to put the data-logger in series (instead of the low value resistor) but the impedance needs to be known. (Previously an ADC-42 was used, this is no longer available and attempts to measure the impedance of the adc-16 have yielded dubious results). I know it’s about 1Mohm but I could do with a few decimal places on that.
Thanks in advance
Rob |
|
| Back to top |
|
 |
RobH Active User

Joined: 01 Nov 2006
Posts: 6
|
Posted: Thu Nov 02, 2006 9:34 am Post subject: |
|
|
I should have pointed out that I need to know the impedance of my ADC-16, not any old generic value.
Thanks  |
|
| Back to top |
|
 |
RobH Active User

Joined: 01 Nov 2006
Posts: 6
|
Posted: Thu Nov 02, 2006 2:53 pm Post subject: |
|
|
Ok, update for you, me, er... (ever get the feeling you're talking to yourself?)
I have used another multimeter (high quality) to measure the impedance of the ADC-16. It worked! (after a fashion).
The problem I have now is that the multimeter outputs 1.01V (as can be seen in the PicoLog software) to make the measurement, this means that I have measured the impedance of the logger's 1V range, does the ADC-16 have multiple ranges? Is it autoranging? (I would imagine so)
If this is the case I can "force" the ADC-16 into the appropriate range by placing a known (very) large resistor in series with the ADC-16 but this makes the measurement a little more... unstable, and it's hard to get an accurate reading.
I hope this simplifies the question, basically I need to know if the ADC-16 has multiple range, and if so, what are they?
Thanks
Rob |
|
| Back to top |
|
 |
Michael Guru

Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire
|
Posted: Mon Nov 06, 2006 9:28 am Post subject: |
|
|
Hello Rob,
Thank you for your posts.
The input range is fixed at ±2.5v, there is no autoranging included.
If you are writing your own software, you will find this information in the product manuals.
Best regards,
Michael |
|
| Back to top |
|
 |
RobH Active User

Joined: 01 Nov 2006
Posts: 6
|
Posted: Mon Nov 06, 2006 9:37 am Post subject: |
|
|
Perfect!
Thanks for your reply Michael.
Best regards
Rob |
|
| Back to top |
|
 |
|