| Author |
Message |
rwalborn Active User

Joined: 30 Jun 2005
Posts: 9
|
Posted: Thu Jun 01, 2006 9:21 pm Post subject: Crest Factor measurement wrong |
|
|
When I measure the crest factor of a sine wave, the crest factor measures 2.865. The crest factor of a sine wave is 1.414. It appears the software is calculating RMS to peak to peak value isntead of RMS to peak.
It is off by a factor of 2.
Pico Scope 2200 series software 5.13.9.
Anyone else seen this? |
|
| Back to top |
|
 |
Michael Guru

Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire
|
Posted: Fri Jun 02, 2006 2:04 pm Post subject: |
|
|
hello,
Thank you for this one. We will look into this for you.
If anyone else has noticed this, please let us know.
Best regards,
Michael |
|
| Back to top |
|
 |
rwalborn Active User

Joined: 30 Jun 2005
Posts: 9
|
Posted: Fri Jun 30, 2006 8:16 pm Post subject: crest factor wrong |
|
|
| Any update on this topic? |
|
| Back to top |
|
 |
Michael Guru

Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire
|
Posted: Wed Jul 05, 2006 2:32 pm Post subject: |
|
|
Hello,
Our software team is working on this.
We should see this in the next software release.
Best regards,
Michael |
|
| Back to top |
|
 |
Michael Guru

Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire
|
Posted: Thu Jul 13, 2006 3:12 pm Post subject: |
|
|
Hello,
The crest factor error has now been fixed.
Please download the latest release of PicoScope 5.14.4 from our software pages.
www.picotech.com/software.html
Best regards,
Michael |
|
| Back to top |
|
 |
rwalborn Active User

Joined: 30 Jun 2005
Posts: 9
|
Posted: Thu Jul 13, 2006 7:19 pm Post subject: Crest factor now fixed. |
|
|
I just downloaded and isntall the latest release and the crest factor is now readign correctly.
Thanks |
|
| Back to top |
|
 |
|