| Author |
Message |
jimmc Newbie
Joined: 15 Feb 2006
Posts: 1
Location: Hampshire U.K.
|
Posted: Wed Feb 15, 2006 11:29 am Post subject: Problem with fast block mode |
|
|
I am trying to use an ADC212 configured as follows:
Recording: Fast Block mode
Input Channels
Timebase: 341uS Number of samples: 32733 (maximum)
Channels: A and B
My problems are:
1 Timestamp seems to run repeatedly from 0 to 4294453 uS and then
start again from 0.
2 Samples from channels A and B share the same timestamp
eg 0, 682,1363, ...
when in fact they are interleaved
eg CHA 0, 682, 1363,...
CHB 341, 1023, 1705, ...
At present I am importing the data into excel and correcting it there.
Am I missing something? Is there a better way?
Jim |
|
| Back to top |
|
 |
Michael Guru

Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire
|
Posted: Wed Feb 22, 2006 9:59 am Post subject: |
|
|
Hello,
The best way to exact actual timings for channel samples is to use the PicoScope software. This will give you 2 channels with independant timing.
You will need to save each block to disk using save on trigger functions.
Best regards,
Michael |
|
| Back to top |
|
 |
|
Powered by phpBB © 2001, 2005 phpBB Group
|