| Author |
Message |
Guest
|
Posted: Fri Jan 06, 2006 8:28 am Post subject: type k and ADC-16 |
|
|
Hi
I am trying to use some type k thermocouples with an ADC-16.
When I set up the software to use the thermocouple, I see a temperature registered, and the temperature rises if I touch the thermocouple. However the temperature is given as very close to 0 deg C. Can you suggest what I am doing wrong? |
|
| Back to top |
|
 |
Michael Guru

Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire
|
Posted: Fri Jan 06, 2006 12:20 pm Post subject: |
|
|
Hello,
The ADC16 is a high resolution(16bit) data logger. However, it does not directly support thermocouples. You will need to amplify the signal.
The AD595 IC is an integrated thermocouple instrumentation amplifier with built-in cold junction compensation. This is not the simplest of ways to measure temperature and great care should be taken with the
circuitry. A lookup table must be used in PicoLog since the AD595 output voltage is not linear with temperature. Further information is available upon request.
Best regards,
Michael |
|
| Back to top |
|
 |
|
Powered by phpBB © 2001, 2005 phpBB Group
|