Posted: Fri Aug 05, 2005 8:23 am Post subject: Problem with timebases in excess of 100mS/div |
|
|
Hope I'm not duplicating anything already written - if so I'm sorry, but I have looked a bit (though perhaps not enough!).
I'm using a Picoscope automotive system, running on a laptop, as supplied by TCD, and find that the waveform display is completely wrong at timebase settings in excess of 100mS. I'm not sure what it it showing, but its not real! Dont have the system at hand at the moment, but the software is probably only about 2 months old, as this is when tha system was purchased |
|
| Back to top |
|
 |
Sarah Guest
|
Posted: Fri Aug 05, 2005 9:32 am Post subject: |
|
|
Hi
Thank you for your post.
I would recommend downloading our latest software and seeing if this helps as there was a bug with these timebases in an earlier version and this could be what you are seeing.
You can download the automotive software from the automotive section of our website.
Best Regards |
|
| Back to top |
|
 |
Autonerdz Zen Master

Joined: 20 Nov 2002
Posts: 208
Location: Washington State USA
|
Posted: Fri Aug 05, 2005 1:56 pm Post subject: Long Timebases |
|
|
| Also be sure you are using Block Mode at the longer time bases. This maintains the sample rate. The other modes will stream at a very slow sample rate, at time bases over 100ms/div, and only very slow signals can be captured accurately. Block mode Will use the internal buffer to capture and then transfer the blocks of data to the PC rather then stream. |
|