Problem with timebases in excess of 100mS/div

 
Post new topic   Reply to topic    Pico Technology Forum Index -> PicoScope
Author Message
Guest






PostPosted: Fri Aug 05, 2005 8:23 am    Post subject: Problem with timebases in excess of 100mS/div Reply with quote

Hope I'm not duplicating anything already written - if so I'm sorry, but I have looked a bit (though perhaps not enough!).

I'm using a Picoscope automotive system, running on a laptop, as supplied by TCD, and find that the waveform display is completely wrong at timebase settings in excess of 100mS. I'm not sure what it it showing, but its not real! Dont have the system at hand at the moment, but the software is probably only about 2 months old, as this is when tha system was purchased
Back to top
Sarah
Guest





PostPosted: Fri Aug 05, 2005 9:32 am    Post subject: Reply with quote

Hi

Thank you for your post.

I would recommend downloading our latest software and seeing if this helps as there was a bug with these timebases in an earlier version and this could be what you are seeing.

You can download the automotive software from the automotive section of our website.

Best Regards
Back to top
Autonerdz
Zen Master
Zen Master


Joined: 20 Nov 2002
Posts: 213
Location: Washington State USA

PostPosted: Fri Aug 05, 2005 1:56 pm    Post subject: Long Timebases Reply with quote

Also be sure you are using Block Mode at the longer time bases. This maintains the sample rate. The other modes will stream at a very slow sample rate, at time bases over 100ms/div, and only very slow signals can be captured accurately. Block mode Will use the internal buffer to capture and then transfer the blocks of data to the PC rather then stream.