Posted: Mon Feb 24, 2003 5:28 pm Post subject: range of allowable samples? |
|
|
Hello,
I am using Picoscope 5.09.04 with an ADC-212/100 on a Win2K PIII laptop. In the menu: settings-scope options-advanced, there is a place to set max samples per scope trace. What is the min & max range that you recomend me to put in this box? At what point do we exede the memory of the ADC-212/100?
Also; please check my math: With max swmples set at 2500, and time scale set at 200microSec/div, the whole window is 2milliSec wide.
so: 2mSec/2500samples = 1/1250000 sec/sample or 1.25Msample/sec
If is correct, then I can always use:
(sample rate) = (max sample)/(screen width) |
|
| Back to top |
|
 |
matthew Site Admin

Joined: 25 Sep 2002
Posts: 133
Location: Cambridgeshire, UK
|
Posted: Fri Feb 28, 2003 12:46 pm Post subject: |
|
|
Rich,
all your calculations are fine, there is nothing hidden so the simple answers are true. The min / max range are inhibited by the dialog box.
The memory of the unit is slightly less than 128,000 samples per channel.
Best Regards, |
|