Posted: Tue Feb 18, 2003 9:54 am Post subject: ADC-212/100 Spectrum usage - noise floor and resolution bw |
|
|
I am from a conventional spec analyser background, using expensive kit to measure simple spectrum emissions. I like the pico concept and want to modernise my tools usage.
I have a few questions regarding the spectrum mode of picoscope, which I have not laid hands on yet, I am eagerly awaiting the first unit.
[1] noise floor - are there any declared limits, for example -120dBm/Hz into a 10kHz res b/w.
[2] after playing with the software demo, how can I work out (correctly) the resolution bandwidth setting over a required measurement frequency range. Is it as simple as dividing the frequency range by the number of points ?
[3] is it possible to limit the start frequencyand stop frequency other than the pull-down toolbar (for the stop frequency only) ?
With thanks
roger34 |
|
| Back to top |
|
 |
matthew Site Admin

Joined: 25 Sep 2002
Posts: 133
Location: Cambridgeshire, UK
|
Posted: Wed Feb 19, 2003 12:42 pm Post subject: |
|
|
Hi Roger,
1. There aren't any specified limits for the noise floor, but the unit's dynamic range approaches 20log(resolution). ie 72dB range for a 12bit ADC-212.
2. Yes, just divide the total frequency range selected by the resolution you have selected from Settings | Options. This will give you the frequency scale increments.
3. Unfortunately not, the Frequency range always starts at 0Hz.
Best Regards, |
|