ADC-16 giving stepped results

 
Post new topic   Reply to topic    Pico Technology Forum Index -> PicoLog
Author Message
Rick Gardner
Newbie


Joined: 28 Oct 2004
Posts: 1

PostPosted: Tue Dec 21, 2004 6:59 am    Post subject: ADC-16 giving stepped results Reply with quote

I am using an ADC-16 with 8 channels. I have set the sampling rate to 5ms single sampling, to match the 200Hz sampling rate stated in the User Manual.

On a channel receiving a smooth input (from an LVDT) the ADC-16 is recording a stepped output, stepping at between 1.14 and 2.3 seconds.

How do I increase the sampling rate?
Back to top
Sarah
Guest





PostPosted: Tue Dec 21, 2004 11:55 am    Post subject: Reply with quote

Hi

Thank you for your post.

The sampling rate of the ADC-16 is not 200Hz, but 1.5Hz. This is split between the active channels. This is why you are seeing stepped results. In order to sample fast you would need a different unit.

Hope this helps

Best Regards