ricardo Moderator

Joined: 07 May 2003
Posts: 98
|
Posted: Mon Dec 15, 2003 11:24 am Post subject: |
|
|
Hi,
I'm sorry you are having problems with the TC-08. Something else you could check is that you don't get overload on one channel, or between channels. The TC08 has an input range of approx 60mV. If more than 5V difference occurs between any of the inputs to the TC08 then incorrect readings will result. Additionally if more than 5V difference occurs between the TC08s ground (also computer ground and usually mains earth) then errors will occur. A typical example where this may occur would be if a thermocouple is used to measure the temperature of the motor on a machine tool. The casing of the motor, although earthed may be at a potential several volts different to that of the earth of the computer.
With regards to the screened thermocouples, we don't sell any but you may find something suitable from Kalestead.
http://www.kalestead.co.uk/
Best regards,
Ric |
|