| Author |
Message |
G.H.Stephenson Guest
|
Posted: Wed Nov 12, 2003 4:03 pm Post subject: Slow data acquisition in Excel |
|
|
Hi, I have setup a DAC16 to run in an excell macro, using all 8 channels, and a userform full of graphs. But for some reason it takes 6 seconds to detect there has been a change by the inputs, and only seems to sample evry 2 seconds. Can you help me?
I can e-mail the program to you if it is any help
Thanks
Graham |
|
| Back to top |
|
 |
markspencer Site Admin

Joined: 07 May 2003
Posts: 610
|
Posted: Thu Nov 13, 2003 8:54 am Post subject: |
|
|
Hi,
I would recommend that you use the function get_cycle, this will let you know when it has a converstion for each of the active channels.
If you are using filtering then this will also increase the amount of time to get a signal.
The ADC-16's sample rate is 1 sample every 1.5 seconds when running at 16 bits this can be sped up if the resolution is reduced.
Best regards, |
|
| Back to top |
|
 |
G.H.Stephenson Guest
|
Posted: Thu Nov 13, 2003 12:29 pm Post subject: |
|
|
Thankyou, I changed to resolution to 14bits, just took a while to figure out why it was giving me wored results, but i have sortred it now,
Thanks |
|
| Back to top |
|
 |
|
Powered by phpBB © 2001, 2005 phpBB Group
|