Labview 8.5 drivers and values not correct?

 
Post new topic   Reply to topic    Pico Technology Forum Index -> LabVIEW
Author Message
GJR
Newbie


Joined: 13 May 2008
Posts: 1

PostPosted: Tue May 13, 2008 11:06 pm    Post subject: Labview 8.5 drivers and values not correct? Reply with quote

Hi
We are running Labview 8.5 and runstream.vi and a picoscope 3424. When we input a DC voltage we don't get the correct reading in the labview code but do get the correct one in the picoscope software.

What we set 1.04 V, 2.01 V, 3.07 V
What we read 1.75 V, 3.32 V 5.06 V

Any Ideas what is going on would really help!
Thanks
Peace
Back to top
gruntman
Guru
Guru


Joined: 28 Sep 2006
Posts: 96
Location: Garden City, KS USA

PostPosted: Thu May 29, 2008 9:16 pm    Post subject: Reply with quote

GJR,

The runstream.vi returns the values from the scope in its raw format. Therefore how are you converting the ADC counts to voltage? The proper way would be the following formula:
(raw * (input_ranges[ch])) / (32766)
Therefore if for example the raw output was 16383 and you had the channel setup on ±10V's, the V reading would be 5.00V's.

Regards,

Richard Boyd
Crag Technologies, Inc
http://www.pc-oscilloscopes.com