PicoScope 6.0.13, trigger level mismatch

 
Post new topic   Reply to topic    Pico Technology Forum Index -> PicoScope R6
Author Message
laichh
Active User
Active User


Joined: 12 May 2008
Posts: 7

PostPosted: Mon May 12, 2008 9:06 am    Post subject: PicoScope 6.0.13, trigger level mismatch Reply with quote

I am using the PicoScope 6.0.13 together with PicoScope 5204.

I am using the supplied 250MHz passive prove & I set the attenuation to x10. At the same time, I also set the "Probe Setting" to x10, so I get exactly what I input into the probe tip.

I then set the trigger to "Rising Edge Trigger".

I do not notice this problem since most of the time, the trigger level is set to 0V. Until this afternoon, I want to trigger to a level other than 0V, I notice that the trigger level do not match my probe setting.

See attachment.

When the probe is set to x10, I wish to trigger at 800mV. As I move the "Trigger Diamond" to 800mV level, the trigger level shows at the bottom of the screen actually indicate 80mV.

I can't figure out a way to make the trigger level follow my probe setting in this case.



PicoScope 5204 Trigger Level Mismatch.PNG
 Description:
This is the screen capture. Note that the "Trigger Diamond" is at 800mV level, but the display at the bottom shows 80mV.

Download
 Filename:  PicoScope 5204 Trigger Level Mismatch.PNG
 Filesize:  68.29 KB
 Downloaded:  273 Time(s)


PicoScope 5204 Setup.JPG
 Description:
This is how I setup an example.

Download
 Filename:  PicoScope 5204 Setup.JPG
 Filesize:  108.13 KB
 Downloaded:  268 Time(s)

Back to top
laichh
Active User
Active User


Joined: 12 May 2008
Posts: 7

PostPosted: Mon May 12, 2008 9:38 am    Post subject: Same problem on PicoScope 6.0.15 Reply with quote

Just updated to 6.0.15, same problem exist.
Back to top
markB
Site Admin
Site Admin


Joined: 27 Mar 2007
Posts: 80
Location: Cambridgeshire,UK

PostPosted: Mon May 12, 2008 12:39 pm    Post subject: Reply with quote

The threshold displayed on the trigger toolbar is the threshold set after your x10 probe has been applied.

When you drop your trigger point onto the scope at 800mV, the inverse of the probe scaling is applied and the scope is configured to trigger at 80mV. It is easy to imagine a probe that could have more than one solution to its inverse scaling; a modulus probe for example. If you were to drop your trigger diamond at 800mV the scope will need to be configured to trigger at both 80mV and -80mV.

Obviously the scope cannot be triggered at more than one threshold and so PicoScope displays the uncaled value to avoid confusion.
Back to top
laichh
Active User
Active User


Joined: 12 May 2008
Posts: 7

PostPosted: Tue May 13, 2008 4:22 am    Post subject: Thank you for your reply Reply with quote

Hi MarkB,

Thank you for your reply. I tried with PicoScope 5.20.1 (with PicoScope 3204) and here is the screen shot.

When I set the probe to x10, & place the "Trigger diamond" to 800mV level, the botton screen shows 800mV (scaled). However, the vertical range at the top of the screen says +/-200mV (Un-scaled).



PicoScope 5.20.1 trigger.PNG
 Description:

Download
 Filename:  PicoScope 5.20.1 trigger.PNG
 Filesize:  65.96 KB
 Downloaded:  220 Time(s)

Back to top
markB
Site Admin
Site Admin


Joined: 27 Mar 2007
Posts: 80
Location: Cambridgeshire,UK

PostPosted: Tue May 13, 2008 7:40 am    Post subject: Reply with quote

You are quite correct. PicoScope 6 does behave differently to PicoScope 5 in this instance. PicoScope 5 cannot create custom probes and as such does not suffer from the problem that I described in my earlier post.
Back to top
laichh
Active User
Active User


Joined: 12 May 2008
Posts: 7

PostPosted: Tue May 13, 2008 8:00 am    Post subject: Will it be aligned? Reply with quote

Hi MarkB,

OK, I understand now. But from user point of view, is kind of mis-leading. Will this "problem" be alligned in future for PicoScope 6 release?
Back to top
markB
Site Admin
Site Admin


Joined: 27 Mar 2007
Posts: 80
Location: Cambridgeshire,UK

PostPosted: Tue May 13, 2008 8:21 am    Post subject: Reply with quote

How would you 'fix' this problem for the modulus probe that I described earlier?

The scope cannot trigger at more than one threshold. It would be even more confusing to have different behaviour for probes that have multiple solutions to their inverse scaling.
Back to top
laichh
Active User
Active User


Joined: 12 May 2008
Posts: 7

PostPosted: Tue May 13, 2008 10:38 am    Post subject: Suggestion Reply with quote

Hi MarkB,

How about we put it this way:

- the trigger threshold level, which displayed at the bottom left of the screen, should always match the vertical scale of the selected trigger source channel.

I understand that the PicoScope 6 allow user to define their custom probe with custom unit, mx+c scalling. If I want to trigger at a specific level of my custom probe (with my custom unit), I should be able to define them easily with the trigger toolbar as well as the "trigger diamond".

Actually, I don't quite understand about the "Modulus Probe" you've mention about, I've never seen any of those.
Back to top
markB
Site Admin
Site Admin


Joined: 27 Mar 2007
Posts: 80
Location: Cambridgeshire,UK

PostPosted: Tue May 13, 2008 10:52 am    Post subject: Reply with quote

A modulus probe would use a look up table to define its behaviour similar to the one below:

-2V 2V
0V 0V
2V 2V

If you input a sine wave into this probe it will give an output like the image below.

Hopefully you can now see that the inverse scaling would yeild more than one trigger point which is impossible for PicoScope.



output.JPG
 Description:

Download
 Filename:  output.JPG
 Filesize:  235.24 KB
 Downloaded:  185 Time(s)

Back to top