Posted: Fri Jan 25, 2008 1:16 am Post subject: Sample size with ADC-212/3 |
|
|
I would like to collect 150,000 sample length records sampled at 2 milliseconds per sample. Can I do this?
The Datalogger software web page states "When used with the ADC-212/3 up to 32 thousand samples per channel can be collected at full speed. At lower speeds up to a million readings can be collected." Is 2 milliseconds low enough to collect the 150,000 samples?
Is there a way to calculate the maximun number of sambles for various speeds?
Thanks |
|
| Back to top |
|
 |
gruntman Guru

Joined: 28 Sep 2006
Posts: 96
Location: Garden City, KS USA
|
Posted: Fri Jan 25, 2008 10:38 pm Post subject: |
|
|
Hello,
With the ADC-212/3 you may use the PicoLog software to achieve what you are wanting to do. You may select a sampling rate up to 1 ms sampling rate and log at this speed for up to 1 million samples.
Regards,
Richard Boyd
Technical Specialist