There is no reason that the thermocouple should be reading incorrectly except under the following conditions:
The thermocouple is faulty, have you tried a different one?
Can you tell me what you are attaching the TC-08 to ... laptop/PC.
If you are attaching it to a laptop it could be noise interference from the mains, as laptops are not usually earthed. The solution to this would be to earth the TC-08 from the 9D Connector to a known and tested grounding point. You should use an earth tester to check the ground.
A possible other reson for these readings is that the oven's ground is not good. Again this should be tested.