| Author |
Message |
br1970 User

Joined: 12 Oct 2007
Posts: 2
|
Posted: Fri Oct 12, 2007 8:41 pm Post subject: 3206 Bandwidth - 200 MHz???? |
|
|
I just had a PicoScope 3206 sent thru our in house metrology department and the resulting bandwidth is around 10 MHz. The test criteria is send a 1.2 Vpp sinewave to the input and find the 3 dB cut off point. I then followed up with my own test at my bench using an Agilent 33220A Function Generator and obtained the following data:
Frequency (MHz) Input (Vpp) Output (Vpp) Db
0.5 1.2 1.198 -0.01448856
1 1.2 1.183 -0.123930028
5 1.2 1.04 -1.242958135
10 1.2 0.8 -3.521825181
20 1.2 0.515 -7.34748034
How do you claim this is a 200 MHz bandwidth scope with these results? Only Channel A was active during these measurements.
Thanks |
|
| Back to top |
|
 |
br1970 User

Joined: 12 Oct 2007
Posts: 2
|
Posted: Fri Oct 12, 2007 10:11 pm Post subject: Found issue |
|
|
| I found the issue after reading the 250 MHz probe specs on your site. I had the probe set to 1:1 which as it states is limited to 10 MHz. Thanks for a great product. |
|
| Back to top |
|
 |
picojohn Site Admin

Joined: 11 Jun 2007
Posts: 361
|
Posted: Mon Oct 15, 2007 1:25 pm Post subject: |
|
|
Dear br1970,
We are delighted when customers are happy!
Regards
John |
|
| Back to top |
|
 |
|