| Hi Dave, I just purchased a Picoscope 3423, I having problem with secondary king lead or plug lead set up, I have low reading or no reading at all with 20:1 attenuator installed. Is there way to fix it or wrong set up. please inform, James, thanks |
|
| Back to top |
|
 |
davidj Site Admin

Joined: 11 Oct 2006
Posts: 67
Location: St. Neots, Cambs, UK
|
Posted: Thu Aug 16, 2007 12:00 pm Post subject: |
|
|
Hi James, The 20:1 attenuator should not be inserted when using either the secondary clamp-on pick up or the COP probe. It would explain the very small signal you are seeing displayed !
Keep the 20:1 attenuator for primary ignition and injector work only. If you use the presets for these measurements you will see (20:1) on the left of the display in the units scale, indicating you need this attenuation for the scale reading to be true.
I hope this helps. |
|