Conecting enviromon el005 with labview

 
Post new topic   Reply to topic    Pico Technology Forum Index -> Getting Started
Author Message
Guest






PostPosted: Mon Jul 23, 2007 2:51 pm    Post subject: Conecting enviromon el005 with labview Reply with quote

Hi. I have to do an application using the logger el005 to acquire data. Running the vi included I dont recieve data from the sensors. DO you have a new VI, that works with it? Or have you detected a problem that may cause it.

It works well using the enviromon.
Back to top
ziko
Site Admin
Site Admin


Joined: 01 Dec 2006
Posts: 606
Location: St Neots

PostPosted: Thu Jul 26, 2007 12:20 pm    Post subject: Reply with quote

Hi and thank you for your post.

What kind of sensors are using, the logger and converter?

Kind regards
Back to top
Guest






PostPosted: Thu Jul 26, 2007 3:52 pm    Post subject: Reply with quote

Hi
I´m using the El005 logger, converter 026 and sensor 030.

Now I can recieve data from them, but just sometimes.
When I run, stop an run again it, sometimes I just recieve 15.4 in both sensors(temp and humidity), but the real values returns when I restart the program.
Is there a close routine that I could be missing? Im using the one included in the example.

Adding some problems, I have seen that the logger dont use the limit values(of sensors, for alarm) of the envimon.ini charged in the labview example, it use the value charged(before) in the envimon with the program button.
Is there a routine, or library that can help me doing this in my oun program?

I hope you can help me

Regards
Abycon
Back to top
gruntman
Guru
Guru


Joined: 28 Sep 2006
Posts: 96
Location: Garden City, KS USA

PostPosted: Wed Sep 19, 2007 7:56 pm    Post subject: Reply with quote

Abycon,

I'm sorry to hear that you are having issues with our example.

I have confirmed your results of 15.42 after the vi is stopped. The way to fix this issue is to delete the close call at the end of the example. There is no need to call the close function as the port is released as soon as the program is closed.

As far as the enviromon.ini you are copying the file from C:\Program Files\Pico Technology\EnviroMon and pasting it into your C:\Program Files\Pico Technology\EnviroMon\Drivers folder aren't you? You can modify this file and the updates will be applied directly to your labview program. Be sure to not have the program running when you modify it and the easiest way to modify it would be to change your settings in EnviroMon program and then program the logger and then copy the file into the drivers folder and rerun your .vi.

Let me know if you have any other questions...

Regards,

Richard Boyd
Crag Technologies, Inc
http://www.pc-oscilloscopes.com
Back to top
Display posts from previous:   
Post new topic   Reply to topic    Pico Technology Forum Index -> Getting Started
All times are GMT
Page 1 of 1

 


Pico Technology topic RSS feed 
Powered by phpBB © 2001, 2005 phpBB Group