Posted: Wed Jun 06, 2007 11:17 am Post subject: |
|
|
Hi and thank you for your post. I have spoken to one of the engineers regarding this.
"Any of the ADC200 range should be adequate for looking at a 2MHz signal.
The most likely explanation of your problem is that the scope probe is not trimmed correctly for the ADC200. You need to re-trim probes when moving them between scopes (or even between channels on the same scope) to compensate for the different input characteristics of each channel. If you are moving the probe between a benchtop scope and the ADC200 he'll need to re-trim it each time."
Hope this helps.
|
|
| Back to top |
|
 |
Jonathan_Jones User

Joined: 06 Jun 2007
Posts: 2
|
Posted: Thu Jun 07, 2007 2:58 pm Post subject: |
|
|
Hi
Thanks for the info - that sorted it
|
|