Data Loggers

How to import calibration data into PicoLog

Products: TC-08, PT-104, RH-02

For this we are going to create a scaling file, this will allow the calibration data to be used for each channel without the need to type everything again.

The folowing data should be saved into a file with an .scl extension, e.g. TC-08cal.scl
The information is to be drawn out of the calibration certificate, including The unit S/N# and calibration details is recommended.

-------------------------------------------------------------------------------------------------

Notes= Calibration data for TC-08 SN# EOR42/330 valid until 11/may/2005

 
Type=Table
Units=°C
-100 -100.07
-50 -49.88
-10 -10.02
0 0.09
50 50.04
100 99.93
200 200.06
400 400.33
600 600.65
800 800.95
1000 1001.31
1200 1201.60

-------------------------------------------------------------------------------------------------

for more information from the help manual, check:
Picolog help (plw.hlp)
index-> scaling -> scaling file

-------------------------------------------------------------------------------------------------

While setting up your measurement
From the window labeled [ "converter xyz" channels ]
Press ‘edit’ which will bring you to the window labeled [ edit "converter xyz" channels ]
Enter the channel name and the sensor type, e.g. thermocouple type, then :

Press on ‘Options’
Press on ‘scaling’
In the “scaling method” field, select ‘Use Scaling from file’
Press on ‘file...’ and select the previously entered scaling file.

Press ok, ok, ok to get back to the [ "converter xyz" channels ] and apply the calibration data to another channel.

Article: KB-153
15 Jun 2007