| Author |
Message |
Guest
|
Posted: Thu Feb 28, 2008 10:34 am Post subject: constant overvoltage showing on 3204 |
|
|
I have just received my new 3204. Running XP SP2 on a Dell Inspiron 9300. I have a supplied probe set to 10x attached to channel A. I cannot get any trace to appear. I have run the demo software and all seems well.
There is a constant red exclamation point showing
! indicating overvoltage for the range selected
Even if I have no probes attached I get this symbol.
I have attached the probe to a standard 12V AC signal using an isolation transformer and nothing shows. I have enabled the 5V P-P 1 KHz square wave and nothing shows.
Do you have any suggestions on this?
Regards,
Geof |
|
| Back to top |
|
 |
picojohn Site Admin

Joined: 11 Jun 2007
Posts: 344
|
Posted: Thu Feb 28, 2008 11:28 am Post subject: |
|
|
Hello Geof,
I am sorry that you are having problems with the PicoScope.
Presuming that the software and device driver are correctly installed, if clicking on the 'Autosetup' icon does not bring the trace into view, then please send an email to support@picotech.com with the following details:
Contact name, address, email and telephone number, with a brief summary of the problem. We will then issue you with a returns number and address for returning the unit for investigation.
Regards
John |
|
| Back to top |
|
 |
Guest
|
Posted: Thu Feb 28, 2008 3:09 pm Post subject: thanks |
|
|
I will retest everything later today. If necessary I will follow thru with an E-mail as you suggested.
Regards,
Geof |
|
| Back to top |
|
 |
Guest
|
Posted: Thu Feb 28, 2008 4:45 pm Post subject: further on constant overvoltage condition |
|
|
One further piece of information. I checked the square wave output of the function generator. It measured 999Hz and a net DC voltage of 2.79V. So it is working fine. I then shifted the probe to Ch B and voila we have a beautiful square wave!!. Shifting the probe back to Ch A just shows the red ! and a flat line at whatever the highest voltage is. Looks to me like I might have a Ch A problem.
A question I have is how long would a repair take with these symptoms? I would like to use the unit for another week on Ch B to diagnose some circuits that I am working on.
Thanks for your assistance.
Regards,
Geof
PS I did register for the forums and this is my user name. |
|
| Back to top |
|
 |
picojohn Site Admin

Joined: 11 Jun 2007
Posts: 344
|
Posted: Fri Feb 29, 2008 11:48 am Post subject: |
|
|
Geof,
Without knowing the exact cause of the fault behaviour, it is impossible to give a meaningful estimate but, we do generally succeed in getting most things turned around within a maximum of two weeks.
If you send the email as requested, we can in the meantime issue an RMA.
Regards
John |
|
| Back to top |
|
 |
Guest
|
Posted: Mon Mar 03, 2008 4:40 pm Post subject: E-mail sent |
|
|
picojohn,
I have sent the E-mail that you requested. Thanks for your assistance.
Regards,
Geof |
|
| Back to top |
|
 |
|