Problem ADC-16 0n Com1

 
Post new topic   Reply to topic    Pico Technology Forum Index -> Serial Port Products
Author Message
domahony
User
User


Joined: 07 Jan 2008
Posts: 2
Location: Dublin

PostPosted: Mon Jan 07, 2008 4:50 pm    Post subject: Problem ADC-16 0n Com1 Reply with quote

I have just experienced a problem with my ADC-16. I set it up on a desk top PC and it works fine. However when I transfered it to a laptop with the same operating system, XP, it could not see it. However when I disconnected the instrumentation, the software found it. As soon as I reconnected the instrumentation it could not find the data logger again. The instrumentation is a set of 4 strain gauges in a full bridge. I'm using the adc-16 to power it. Is there a problem with this and do I need to use an external power supply?

Dave
Back to top
picojohn
Site Admin
Site Admin


Joined: 11 Jun 2007
Posts: 311

PostPosted: Wed Jan 09, 2008 3:11 pm    Post subject: Reply with quote

Hello Dave,

The amount of current that may be drawn from Pin 12 on the ADC-11 varies between computers. It is very likely that your laptop PC is unable to supply the same level of current as your desk top PC, and therefore, too much current is drawn from pin 12, leading to an excessive drop in the ADC-11 supply voltage from the PC.

As you have already deduced, the solution would be an external supply for your instrumentation.

Regards
John
Back to top
domahony
User
User


Joined: 07 Jan 2008
Posts: 2
Location: Dublin

PostPosted: Thu Jan 10, 2008 12:36 pm    Post subject: Reply with quote

Cheers John,
Solved by putting on a battery whihc for the tests works well.

Dave
Back to top