Hi:

When I use "Freeze/thaw one" and composite tool I can´t see the captured traces at the same screen.
I am working with single mode and only a channel.

Could you help me?

I use 5.16.12 PS ver and a adc212/50 and a 3204 too

Thank you and regards,

Ariel
Back to top
ziko
Site Admin
Site Admin


Joined: 01 Dec 2006
Posts: 592
Location: St Neots

PostPosted: Fri Dec 28, 2007 2:28 am    Post subject: Reply with quote

Hi Ariel,

To get it to work you need to do the following:

Open two scope screens "View->New Scope",
Then select the "View->Freeze/Thaw one" on the desired screen
Finally select "View->New Composite "

Hope this helps.

Kind regards
Back to top
arielmedina61
Super User
Super User


Joined: 23 Sep 2005
Posts: 20
Location: Buenos Aires-Argentina

PostPosted: Fri Dec 28, 2007 11:49 am    Post subject: Reply with quote

Hi,Ziko

Thank you for your reply.

I know how this tool works, but when I capture one trace on the 2nd screen, the first one disappear and the composite screen shows only one trace. Remember I use single trigger mode. I think this tool does not work with this trigg mode.


Thank you and regards,


Ariel
Back to top
arielmedina61
Super User
Super User


Joined: 23 Sep 2005
Posts: 20
Location: Buenos Aires-Argentina

PostPosted: Thu Jan 03, 2008 1:30 pm    Post subject: Reply with quote

Hi Ziko

I am waiting for your reply!

Good year!


Ariel
Back to top
arielmedina61
Super User
Super User


Joined: 23 Sep 2005
Posts: 20
Location: Buenos Aires-Argentina

PostPosted: Tue Jan 08, 2008 12:11 pm    Post subject: Reply with quote

Hi

Any help would be very grateful!!


Thaks

Ariel
Back to top
markspencer
Site Admin
Site Admin


Joined: 07 May 2003
Posts: 610

PostPosted: Wed Jan 09, 2008 9:29 am    Post subject: Reply with quote

Hi Ariel,

I am sorry to hear that you are experiencing this problem. I have tried to the freeze/thaw functionality with the composite view without incident

The version of the software I was using was Release 5.18.0. I noticed that you where using an older version of the software Release 5.16.12, I would recommend that you upgrade to the newer version and then try the freeze/thaw and composite function again.

The software can be downloaded from:

http://www.picotech.com/download.html
Back to top
arielmedina61
Super User
Super User


Joined: 23 Sep 2005
Posts: 20
Location: Buenos Aires-Argentina

PostPosted: Thu Jan 10, 2008 6:07 pm    Post subject: Reply with quote

Hi Mark,

Thank you for your reply.

Following your advice, I ´ve downloaded the new version 5.18.0.

However there is a problem when I use single trigger. If I use AUTO there is no problem. Also I reduced the number of samples but the the result is the same.

Changing the trigger from single to auto and vice versa the scope re start and it can capture a new trace. Then I open a Composite view and works OK.

I going to use this procedure to capture differents traces to make a comparison.
Regards,
Ariel.