| Author |
Message |
Guest
|
Posted: Thu Sep 27, 2007 5:24 am Post subject: tc-08 |
|
|
| im using 3 channels for temp & 1 channel for pressure through a transducer what is required help im STUCK.thank you |
|
| Back to top |
|
 |
ziko Site Admin

Joined: 01 Dec 2006
Posts: 586
Location: St Neots
|
Posted: Thu Sep 27, 2007 9:13 am Post subject: |
|
|
Hi and thank you for your post.
Are you using the USB TC-08 or the serial TC-08. In any case the USBTC-08 has +/-60mV range on all the channels whilst the USB TC-08 has +/-70mV.
As long as the output of the pressure transducer is within the range of the TC-08 then you can measure it. Then using our software you can use the scaling option to convert the readings to your desired measurement.
Also in the software settings when selecting each channel, you have the option to choose all the various kinds of thermocouples. At the end of the selection you also have a mV settings, choose this for your pressure measurement.
Let me know how you get on, or if you have any further questions please do not hesitate to contact me.
Kind regards
Ziko |
|
| Back to top |
|
 |
Guest
|
Posted: Wed Oct 03, 2007 3:43 pm Post subject: TC-08 + pressure transducer |
|
|
| Do you have suggestions - places to purchase the pressure transducer to use with the TC-08? |
|
| Back to top |
|
 |
ziko Site Admin

Joined: 01 Dec 2006
Posts: 586
Location: St Neots
|
|
| Back to top |
|
 |
|