Posted: Tue Jul 17, 2007 1:03 am Post subject: TC-08 Temp adjustment |
|
|
It's been a while since i worked with this unit. I seem to recall a way to adjust / calibrate temperatue differences a few degrees up or down.
Like 3 t/c's reading 88, 86, 88 being able to adjust the 86 up to 88.
Can this be done?
Thanks,
David |
|
| Back to top |
|
 |
ziko Site Admin

Joined: 01 Dec 2006
Posts: 586
Location: St Neots
|
Posted: Tue Jul 17, 2007 10:01 am Post subject: |
|
|
Hi and thank you for your post.
Using PicoLog you can use a look up table when you select your input channels, you would need to do the following:
Options -> Scaling-> Scaling Method (Look Up Table)
you would then for example put the following:
0 4
2 6
5 9
....
86 88
So you would have the raw value then a space then the scaled value.
Hope this helps.
Kind regards |
|
| Back to top |
|
 |
DGCYS User

Joined: 17 Jul 2007
Posts: 3
Location: Dom. Rep.
|
Posted: Wed Jul 18, 2007 12:45 am Post subject: Thanks |
|
|
That helps but after applying a filtered value of 20 all 6 stabilzed to within 1ยบ of each other. This is very acceptable considering i made the T/C's 60 feet long
Thanks,
DAVID |
|