Andrei Guest
|
Posted: Thu Sep 14, 2006 8:41 pm Post subject: PT-104: precision when measuring 0.5 Ohm |
|
|
I want to measure specimens within resistance range approx. 0.5 ... 200 Ohm with precision of 0.25%. It looks like PT-104 is up to the task, but I am not quite sure about the measurement precision in the lower range.
Can you tell me what is the relative precision of resistance measurement at 0.5 Ohm, please?
regards,
Andrei |
|
| Back to top |
|
 |
Michael Guru

Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire
|
Posted: Mon Sep 18, 2006 8:27 am Post subject: |
|
|
Hello Andrei,
The accuracy of resistance measurement is to within 1 µΩ when using the 0-375Ω range. This will cover the range you wish to measure.
Best regards,
Michael |
|
| Back to top |
|
 |
Andrei Guest
|
Posted: Mon Sep 18, 2006 11:06 am Post subject: |
|
|
I am no expert on measurement theory, thus, please correct if I am wrong.
1uOhm looks more like a resolution.
RMS noise on range 375Ohm is 10 ppm, which makes 3.75 milliohm, i.e. accuracy 0.75% when measuring 0.5Ohm.
Thus, if I want to lower RMS noise to 0.25%, I have to average over (0.75/0.25)^2 = 9 measurements, which is not that bad. |
|