TC-08 uncertainty

 
Post new topic   Reply to topic    Pico Technology Forum Index -> Pre-Sales Advice
Author Message
Dan Acres
Guest





PostPosted: Wed Sep 13, 2006 12:18 pm    Post subject: TC-08 uncertainty Reply with quote

Hello, I'm looking at the TC-08 for a high accuracy temperture monitoring system within an enclosure.
I'm not quite clear on what the uncertainty on the TC-08 actually is in high accuracy resolution (0.025K) mode.
According to the online spec the uncertainty in measurement is approx. 1 Kelvin at room temperature. Is this correct?
Back to top
Sarah
Guest





PostPosted: Wed Sep 13, 2006 12:23 pm    Post subject: Reply with quote

Hi Dan

The accuracy of the TC-08 is the sum of ±0.2% and ±0.5C. The first refers to the measurement and the second to the cold junction measurement. The accuracy of the sensor also needs to be taken into account.

Therefore at room temperature the accuracy would be less than 1C.

Best Regards
Back to top
Guest






PostPosted: Wed Sep 13, 2006 12:28 pm    Post subject: Reply with quote

Sarah wrote:

The first refers to the measurement


In centigrade instead of an absolute measurement?
Back to top
Sarah
Guest





PostPosted: Wed Sep 13, 2006 12:29 pm    Post subject: Reply with quote

Hi

Yes it is with reference to centigrade.

Best Regards
Back to top
Dan Acres
Guest





PostPosted: Wed Sep 13, 2006 12:31 pm    Post subject: Reply with quote

ok, cheers
Back to top