PostPosted: Wed Sep 06, 2006 8:36 pm    Post subject: 3206 sensitivity Reply with quote

I am seriously considering buying a 3206 kit. However, the technical specification states a vertical sensitivity of +/- 100mV to +/- 20V/div. Elsewhere the sensitivity is stated as 10mV to 2V/div. Which is correct?
Also, the stated resolution is 8 bits - are 8 bits used at 100mV and also at 20V?
Back to top
Michael
Guru
Guru


Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire

PostPosted: Thu Sep 07, 2006 8:30 am    Post subject: Reply with quote

Hello,

To clarify the resolution, I will show you how to calculate A-D converter steps.

The calculation goes like this with only two variables to change.
A-D mV steps = Voltage range ÷ Bit resolution

Now, you will need to compute the bit resolution like this: 2x10^n where n=bit resolution. Next you will need to compute the voltage range. A ±500mV range covers -500mV to +500mV, a range of 1000mV.

So the calculation now looks like this:
3.9mV = 1000mV ÷ 256 (2x10^8)

Hope this makes sense.

Best regards,
Michael
Back to top
Carl Grotti
Guru
Guru


Joined: 21 Nov 2002
Posts: 50
Location: Missouri, USA

PostPosted: Thu Sep 07, 2006 6:19 pm    Post subject: Reply with quote

Hello Michael,

I hope you don't mind me jumping in here.

Your step calculation makes perfect sense except for one thing. How does 2x10^8 =256? I'm not familiar with the "^" symbol.

Thanks,
Back to top
Michael
Guru
Guru


Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire

PostPosted: Fri Sep 08, 2006 9:42 am    Post subject: Reply with quote

Hello Carl,

Sorry for my lack of ASCII codes, I was trying to represent 2 power 8 in symbol and used the carat symbol for lack of superscripts.

The calculator version would be like this: (2^8)
3.9mV = 1000mV ÷ 256 (2^8)

Sorry for the error.

Regards,
Michael
Back to top
rayblane
Guest





PostPosted: Tue Sep 12, 2006 9:35 am    Post subject: 3206 resolution Reply with quote

Thanks for the replies. You have answered the question about resolution, but not about sensitivity.
Which of the two quoted vertical deflection ranges is correct, 10mV to 2V or 100mV to 20V?

Thanks.
Back to top
Sarah
Guest





PostPosted: Wed Sep 13, 2006 11:21 am    Post subject: Reply with quote

Hi

Thank you for your post.

The correct ranges are from 100mV to 20V - the maximum input is ±20V.

Best Regards