Posted: Wed Jun 07, 2006 5:03 am Post subject: Average calculation fails sometimes |
|
|
Under certain circumstances a measurement with average-calculate fails.
Please try the following simple test-setup:
1. Signal-Generator: 1 kHz square, Sweep Mode active, Repeat, Stop frq 2 kHz, Increment 1.0.
2. chA: input from signal-generator, 200 us, Trigger chA Auto.
3. Settings/Measurement: chA, frequency, whole trace.
4. Settings/Measurement/Options: Statistics, Current Value, Average, Standard deviation, No of readings= 10000
Now - this is important - do the following: Settings/Options/OK (with or without doing any changes)
Start recording and after about 3 seconds stop recording
You can see that the average and std deviation calculation doesn't stop, it runs until the reading count is reached. Even if you remove the input from chA the calculation will not stop. Of course the values are not correct anymore after the stop.
But there is hope: repeat this experiment with Settings/Options/CANCEL - now it works as expected: average calculation stops - also when removing the input signal from chA.
You can repeat this error On/Off scenario as often as you like.
Best regards, Karl |
|
| Back to top |
|
 |
Guest
|
Posted: Wed Jun 07, 2006 11:26 pm Post subject: |
|
|
I also see this problem often,
Restarting Picoscope fixes it, for awhile. |
|
| Back to top |
|
 |
Guest
|
Posted: Thu Jun 08, 2006 7:55 am Post subject: |
|
|
No reason for a restart - just use the sequence 'Settings/Options/Cancel'
Force the error with: 'Settings/Options/OK' |
|
| Back to top |
|
 |
Michael Guru

Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire
|
Posted: Thu Jun 08, 2006 8:20 am Post subject: |
|
|
Hello,
All your posts are being considered here. Please let us know if you are experiencing any other problems.
Best regards,
Michael |
|