ADC-20 Sample time

 
Post new topic   Reply to topic    Pico Technology Forum Index -> USB PC Data Loggers
Author Message
Shaun
User
User


Joined: 18 Jan 2006
Posts: 2

PostPosted: Wed Jan 18, 2006 9:00 pm    Post subject: ADC-20 Sample time Reply with quote

Hi.

I'm using the ADC-20 in a VB application. To sample a single channel requires 60mS which seems quite long. Is this a hardware limitation or a software driver limitation ?

Thanks
Back to top
Michael
Guru
Guru


Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire

PostPosted: Thu Jan 19, 2006 3:59 pm    Post subject: Reply with quote

Hello Shaun,

The limitation is down to the resolution of the measurement which equates to 1,048,576 bits. Running the ADC20 at 60ms conversion times reduces the resolution but runs the ADC20 faster.

Best regards,
Michael
Back to top
RobH
Active User
Active User


Joined: 01 Nov 2006
Posts: 6

PostPosted: Wed Nov 08, 2006 11:24 am    Post subject: Reply with quote

This is going to sound like a bit of a dumb question Embarassed ... but here goes nothing:

I am using an ADC-16 to log voltage, I am measuring only a few milli volts and hence the 16 bit accuracy is nice to have, but I would also like to sample at 50-100ms intervals. Whilst I appreciate there is no way to increase the conversion time (~600ms), is it worth sampling at 50ms intervals? Might I "get lucky" every so often as recieve more frequent value updates?

Thanks in advance Wink

Rob
Back to top