Posted: Wed Oct 26, 2005 4:52 pm Post subject: 3206 and 10GS/s sampling rate for repetitive signals??? |
|
|
Hello, I have a question - what does the 10GS/s sampling rate of this oscilloscope exactly mean? For what can this feature be used? I would like to visualize eye diagrams of modern fast serial buses, like PCIexpress, gigabit Ethernet or Infiniband (1 ~ 2.5 Gbps). Aren't there low pass filters at input of the ADC, which would make it impossible to measure such signals? Can this feature be used to for oversampling?
Thanks
Gordon |
|
| Back to top |
|
 |
Michael Guru

Joined: 07 Jul 2005
Posts: 93
Location: St Neots, Cambridgeshire
|
Posted: Tue Nov 01, 2005 4:25 pm Post subject: |
|
|
Yes it can Gordon.
The software allows you to select ETS functions and oversampling rates for the 3204,5&6.
Best regards,
Michael |
|
| Back to top |
|
 |
Sarah Guest
|
Posted: Tue Nov 08, 2005 9:38 am Post subject: |
|
|
Hi
Just to add to that, the front end bandwidth is 200MHz on the 3206 and so anything over this may not display as well.