PicoScope 4224 IEPE Oscilloscope Specifications
Listed here are the specifications for the PicoScope 4224 IEPE. This advanced version of the PicoScope 4224 features an IEPE interface making it compatible with industry-standard IEPE accelerometers and microphones.
| Oscilloscope — Vertical | ||
|---|---|---|
| Mode | Passive Probe Mode | IEPE Interface Mode |
| Bandwidth | DC to 20 MHz (10 MHz on ±50 mV range) | 1.6 Hz to 20 MHz (10 MHz on ±50 mV range) |
| Rise time (calculated) | 17.5 ns on all ranges except 50 mV; 35 ns on 50 mV range | |
| Output | not applicable | 4 mA up to 24 V ±5% |
| Input channels | 2 | |
| Vertical resolution | 12 bits | |
| Enhanced vertical resolution | 16 bits | |
| DC accuracy | ±1% of full scale | |
| Input sensitivity | 10 mV/div to 4 V/div | |
| Input characteristics | 1 MΩ in parallel with 22 pF | 1 MΩ in parallel with 1 nF |
| Input type | Single–ended, BNC connector | |
| Input coupling | Software selectable AC/DC | AC |
| Input ranges (full scale) | ±50 mV to ±20 V in 9 ranges | |
| Overload protection | ±100 V | |
| Oscilloscope — Horizontal | ||
|---|---|---|
| Maximum sampling rate (single shot) | 80 MS/s | |
| Maximum sampling rate (continuous streaming mode) | 1 MS/s (Record length limited to 20 MS in PicoScope, unlimited when using the supplied SDK) | |
| Buffer memory | 32 M samples shared between active channels | |
| Waveform buffer | Up to 10000 waveforms | |
| Timebase ranges | 100 ns/div to 200 s/div | |
| Timebase accuracy | 50 ppm with ≤10 ps jitter | |
| Triggers | ||
|---|---|---|
| Source | Any input channel | |
| Basic triggers | Auto, repeat, single, rapid, none | |
| Advanced triggers | Rising edge, falling edge, edge with hysteresis, pulse width, runt pulse, dropout, windowed | |
| Maximum pre trigger delay | 10 x “per division” timebase setting | |
| Maximum post trigger delay | 50 s | |
| Spectrum Analyser | ||
|---|---|---|
| Bandwidth | 20 MHz | |
| Frequency range | DC to 20 MHz | |
| Maximum number of bins | 1,048,576 | |
| Display modes | Magnitude, peak hold, average | |
| Window types | Blackman, Gaussian, triangular, Hamming, Hann, Blackman–Harris, flat–top, rectangular | |
| PC Requirements | ||
|---|---|---|
| Minimum | Processor: Pentium II processor, or equivalent Memory: 64 MiB (XP) / 512 MiB (Vista) / 1 GiB (Win 7) Operating system: 32– or 64–bit edition of Microsoft Windows XP SP2, Vista or Windows 7 Ports: USB 2.0 compliant port | |
| Recommended | Processor: 2 GHz Pentium IV processor, or equivalent Memory: 256 MiB (XP) / 1 GiB (Vista / Win 7) Operating system: 32– or 64–bit edition of Microsoft Windows XP SP2 (or above), Vista or Windows 7 Ports: USB 2.0 compliant port | |
| Environmental | ||
|---|---|---|
| Operating environment Temperature range Humidity | 0 °C to 45 °C (20 °C to 30 °C for quoted accuracy) 5% to 80% RH, non-condensing | |
| Storage environment Temperature range Humidity | -20 to +60°C 5 to 95% RH, non-condensing | |
| Physical Properties | ||
|---|---|---|
| Dimensions | 200 x 140 x 35 mm (approx 7.9 x 5.5 x 1.4 in) | |
| Weight | <500 g (approx 1.1 lb) | |
| Software | ||
|---|---|---|
| PicoScope 6 for Windows | PicoScope 6 is your complete test and measurement lab in one application.
Features include: Capture modes: oscilloscope, spectrum and persistence. Channel maths: calculate the sum, difference, product, inverse or create your own custom function using standard arithmetic, exponential and trigonometric functions. Mask limiting testing: pass/fail, failure count, total count. Serial Decoding: decode data from a serial bus such as I²C. Automated measurements Scope mode: AC RMS, true RMS, cycle time, DC average, duty cycle, falling rate, fall time, frequency, high pulse width, low pulse width, maximum, minimum, peak–to–peak, rise time and rising rate. Spectrum mode: frequency at peak, amplitude at peak, average amplitude at peak, total power, total harmonic distortion (THD % and THD dB), total harmonic distortion plus noise (THD+N), spurious-free dynamic range (SFDR), signal+noise+distortion to signal+noise ratio (SINAD), signal to noise ratio (SNR) and intermodulation distortion (IMD). Export data formats: comma separated values (CSV), tab delimited (TXT), windows bitmap (BMP), graphics interchange format (GIF), portable network graphics (PNG), MATLAB 4 format (MAT). Full details for PicoScope 6 | |
| Software development kit | Includes drivers and example code for various programming languages including C, Excel, LabVIEW and Visual Basic. Download SDK | |
| Language Support | ||
|---|---|---|
| Software PicoScope 6 | Full support for: English, Français, Deutsch, Italiano, Español Menus and dialogs only for: 中文 (简体), 正體中文 (繁體), Čeština, Dansk, suomi, Ελληνικά, Magyar, 한국어, 日本語, Norsk, Polski, Português, română, Русский, Svenska, Türkçe | |
| Documentation User’s guide Programmer’s guide | English, Français, Deutsch, Italiano, Español, 中文 English | |
| General | ||
|---|---|---|
| Data sheet | English, 中文 (简体), Español, Français, Italiano | |
| Additional hardware (supplied) | USB 2.0 cable, user manuals, software CD–ROM | |
| PC interface | USB 2.0 (USB 1.1 compatible) | |
| Power requirements | Powered from USB port | |
| Compliance | European EMC and LVD standards FCC Rules Part 15 Class A RoHS compliant | |
| Total Satisfaction Guarantee | In the event that this product does not fully meet your requirements you can return it for an exchange or refund. To claim, the product must be returned in good condition within 14 days. | |
| Warranty | 5 years | |
| Lead time | In stock | |
| Price | £599·00 US$988·35* €724,79* | |
* US Dollar and Euro prices are approximate.
Note: quoted specifications apply when using PicoScope 6 software.
Need to know a specification not listed here? — just ask
Cart
Share