PicoScope 3425 PC Oscilloscope Specifications
| Oscilloscope | |
|---|---|
| Bandwidth | 5 MHz (3 MHz on 100 mV range) |
| Channels | 4 |
| Vertical resolution | 12 bits |
| Enhanced vertical resolution | 16 bits |
| Maximum sampling rate (single shot) One channel in use Two channels in use Three or four channels in use | 20 MS/s 10 MS/s 5 MS/s |
| Record length One channel enabled, up to 100 ms/div Two channels enabled, up to 100 ms/div Three or four channels enabled, up to 100 ms/div Over 100 ms/div | 512 kS/channel 256 kS/channel 128 kS/channel 2 MS/channel |
| Waveform buffer | Up to 1000 waveforms |
| Input impedance | 12.4 MΩ (on 100 mV to 5 V ranges) 10.1 MΩ (on 10 V to 400 V ranges) |
| Input capacitance | 12 pF |
| Input type | Differential voltage, Touch-proof BNC connectors with 4 mm GND sockets |
| Input coupling | Software selectable AC/DC |
| Input voltage ranges | 100 mV to 400 V in 12 ranges |
| Common-mode voltage range To ensure measurement accuracy | 30 V (on 100 mV to 5 V ranges) 400 V (on 10V to 400V ranges) |
| Maximum safe voltages Differential Any input above scope GND | 400 V 400 V (600 V transient) |
| Measurement category rating | CAT 1 |
| Accuracy Voltage Time | ±1 % 50 ppm |
| Linearity | 12 bits |
| Noise | < 10 LSB |
| PC Requirements | |
|---|---|
| Minimum | Processor: Pentium II processor, or equivalent Memory: 64 MB (XP) / 512 MB (Vista) minimum Operating system: Microsoft Windows XP SP2 (32-bit) or Vista (32 or 64-bit) Ports: USB 1.1 compliant port |
| Recommended | Processor: 2 GHz Pentium IV processor, or equivalent Memory: 256 MB (XP) / 1 GB (Vista) Operating system: Microsoft Windows XP SP2 (32-bit) or Vista (32 or 64-bit) Ports: USB 2.0 compliant port |
| Environmental | |
|---|---|
| Operating temperature range | 0 °C to 40 °C ( 20 °C to 30 °C for quoted accuracy) |
| Operating humidity range | Minimum 5% RH non-condensing Maximum 80% RH non-condensing, decreasing linearly to 50% at 40 °C |
| Storage environment Temperature range Humidity | -20 °C to +60 °C 5% to 90% RH, non-condensing |
| Other environmental conditions | Dry environments Altitudes of up to 2000 m above sea level No pollution, or only dry, non-conductive pollution |
| Physical Dimensions | |
|---|---|
| Dimensions | 255 x 170 x 40 mm (approx 10.0 x 6.7 x 1.6 in) |
| Weight | 920 g (approx 2 lb) |
| Software | ||
|---|---|---|
| PicoScope 6 for Windows | PicoScope 6 is your complete test and measurement lab in one application. Features include: Capture modes - oscilloscope, spectrum and persistence modes Channel maths - calculate the sum, dfference, product, inverse or create your own custom function using standard arithmetic, exponential and trigonometric functions Automated measurements Scope mode: AC RMS, cycle time, DC average, duty cycle, falling rate, fall time, frequency, high pulse width, low pulse width, maximum, minimum, peak-to-peak, rise time and rising rate Spectrum mode: frequency at peak, amplitude at peak, total power, total harmonic distortion (THD), total harmonic distortion plus noise (THD+N), spurious-free dynamic range (SFDR), signal+noise+distortion to signal+noise ratio (SINAD), signal to noise ratio (SNR) and intermodulation distortion (IMD) Export data formats - Comma separated values (CSV), tab delimited (TXT), windows bitmap (BMP), graphics interchange format (GIF), portable network graphics (PNG), MATLAB 4 format (MAT) Full details on PicoScope 6 | |
| PicoLog for Windows | PicoLog data acqusition software can collect up to 1 million samples. Features include: Multiple views - view data as a graph, spreadsheet or text Parameter scaling - convert raw data into standard engineering units Math functions - use mathematical equations to calculate additional parameters Alarm limits - program an alert if a parameter goes out of a specified range IP networking - transfer measurements via a LAN or over the Internet Full details on PicoLog | |
| Linux drivers | 32-bit and 64-bit | |
| Software development kit | A growing collection of drivers and example code for various programming languages | |
| Language support | |
|---|---|
| Software PicoLog PicoScope 6 | Full support for English, French and German; menus only for Italian and Swedish Full support for English, German and Spanish; menus only for French, Italian, Dutch, Hungarian and Simplified Chinese |
| Documentation User’s guide Installation guide | English English, French, German, Italian, Spanish |
| General | |
|---|---|
| Additional hardware (supplied) | Carry case USB 2.0 cable User manuals and software CD-ROM 2 x test probes 4 x screened BNC to 4 mm plug leads 4 x current clamp adaptors 4 x small crocodile clips 4 x large crocodile clips 8 x sprung hook probes |
| PC interface | USB 2.0 (USB 1.1 compatible) - cable supplied |
| Power requirements | Powered from USB port |
| Compliance | FCC, CE RoHS compliant |
| Total Satisfaction Guarantee | In the event that this product does not fully meet your requirements you can return it for an exchange or refund. To claim, the product must be returned in good condition within 14 days. |
| Warranty | 2 years |
| Price * US Dollar and Euro prices are approximate. | £975.00 $1608.75* €1218.75* |
Note: quoted specifications apply when using PicoScope 6 software.
Need to know a specification not listed here? - just ask
Cart
Share