Here we will be modelling the Amplitude ultrasound scan or A-scan used in ophthalmology. It is the simplest type of scanning, relying on a one-dimensional pulse-echo technique similar to that used in the echo location (SONAR) of fish.

Whilst this analogy with ultrasound A-scanning is good, you will need to emphasise that the two coaxial cables of differing electrical impedance are representing two different materials in the eye of differing acoustic impedance. By knowing the speed of the pulses and where reflections occur, they can then calculate the lengths of the cables in a similar way to which distances to parts of the eye would be determined. With ultrasound there are far less multiple reflections of note, although false echoes can occur between tissue and bone, tissue and air, and tissue and the transducer. In most cases there would also be appreciable scattering and so the amplitude of any multiple reflections would prove tiny. With the electrical pulses the students should be able to see the multiple reflections as there is less scattering and attenuation.

If students are shown the original output pulse they will then note that, on reflection at the cable junctions, there is a phase change with the pulse being inverted. This is of no consequence with the ultrasound scan as it is just the amplitude of the reflected pulse that is important.

As students will have met propagation of light by internal reflection down optical fibres, it may be wise to comment that the electrical pulses effectively travel in a straight line down the cables as if they were laid out straight, there is no ‘bouncing’ off of the walls of the cables. They may well be surprised at the speeds of pulses in such cables being a substantial fraction of the speed of light.

The activity, whilst modeling the ultrasound A-scan, is a useful one for seeing how to obtain times from an oscilloscope screen and, in general, how to set up and use an oscilloscope. Whilst the activity could be conducted on a traditional oscilloscope (ideally of the storage type) it is certainly easy to do with the PicoScope 2202. If you have the older ADC-200 version then the same activity can be conducted with it too, as indeed it could with a PicoScope 3000 series device. Obviously it does need an oscilloscope with a fast timebase.

The speeds of the pulses down coaxial cables tends to be around 2 x 108 m/s. The two I used gave speeds of 1.95 x 108 m/s and 1.99 x 108 m/s which are rather close to each other. As the real speeds are not important I have suggested stating a speed of 1.90 x 108 m/s for one cable and 2.10 x 108 m/s for the other. The lengths they calculate will then not actually be the real ones, but there again I suspect no one is going to wish to unwind them and wind them back on the reels again, so that does not matter.

Typical results

Output pulse: 708ns. First main reflection: 1681 ns. Second main reflection: 3736 ns.

Time for pulse to travel to end of cable A and back: (1681 - 708) ns = 973 ns. Speed of pulse down cable A is 2.10 x 108 ms-1. So length of cable A is given by:

Speed of pulse x ½ x Time for pulse to travel = (2.10 x 108) x 0.5 x (973 x 10-9) = 102 m

Time for pulse to travel through cable A to the end of cable B and back: (3736 - 708) ns = 3028 ns. However, 973 ns is the time that this pulse was traveling through cable A, so the time in cable B must be (3028 - 973) ns = 2055 ns. Speed of pulse down cable B is 1.90 x 108 ms-1. So length of cable B is given by:

Speed of pulse x ½ x Time for pulse to travel = (1.90 x 108) x 0.5 x (2055 x 10-9) = 195 m

Answers to questions

Q1 Using the expression Ir/Ii = (Z2 - Z1)2/(Z2 + Z1)2 and substituting for Z1 = 1.38 x 106 kg m-2 s-1 and Z2 = 6.5 x 106 kg m-2 s-1 we then have:

Ir/Ii = (6.5 x 106 0 1.38 x 106)2/(6.5 x 106 + 1.38 x 106)2= 0.42 (42%)

Q2 Answers will depend on the results the students got but, assuming the cables are very similar to those I recommended cable A should be near 100 m in length and cable B near 200 m in length.

Useful websites

A-scan Biometry
Whilst this is a technical website for use by the medical profession, it does provide a fair bit of detail about the process and some images.

Medical Physics Teaching Materials for Schools
This University College London website gives access to an enormous range of resources for medical physics teaching. A first stop for such including Powerpoint presentations with superb imagery.

The Principles of Medical Ultrasound
This is an excellent site outlining the processes employed, how the ultrasound is produced, what acoustic impedance is and why it is important, together with a little on the B-scan, advantages and disadvantages of ultrasound scanning, and Doppler ultrasound scanning. Well worth looking at.

Types of Ultrasound
This Anaesthesia Uk website displays the many types of ultrasound scanning used in medicine, supported by excellent images — some as movies.

Useful books

Health Physics. A K McCormick and A T Elliot. Cambridge University Press. 1996.
ISBN 0 521 42155 1

Medical Physics. Martin Hollins. Macmillan Education. 1990.
ISBN 0 333 46657 8

Physics in the Life Sciences. George Duncan. Blackwell Scientific Publications. 1990.
ISBN 0 632 01778 3

Scientific Basis of Medical Imaging. Editor P N T Wells. Churchill Livingstone. 1982.
ISBN 0 443 01986 X

NOTE: The last of these is a highly technical publication and would almost certainly have to be requested for loan from a library. It is not one to purchase! It has, however, all the detail one might ever want on ultrasound and other forms of scanning used in medicine.